sentinel
 
vrx
 
sentinel
 
VR HD
 
sentinel
 
sentinel
 
VRx Dive computers Sentinel VR Technology Home
Ouroboros Rebreathers by VR Technology
more about our unit| instructors and training courses| Technical Support Services| Technical specification of our engineering| Updates and new downloads| Read about our history and see some old pictures| contact us|photos of the unit | Join in on the Ouroboros mailing forum| Find out who we are and what we are about |Central navigation page
Ouroboros rebreather Oxygen Cells
note; Milli volts of each cell can be tested using a button on the passive display.
Oxygen Cells
The Ouroboros rebreather uses R-17D oxygen cells with mini jack connectors which are located within the center section of the rebreather unit. The location is fundamental to the health of the cells and is designed in such a way that they have limited exposure to moisture across the membranes thus avoiding problematic cell function during dives. The mini jack connectors enable the user to simply unplug the cell block and remove it completely from the unit, no extra tools are required.

Read on or jump to one of the menus below

Purchase cells

Computer Sensor screens
Sensor Calibration
Testing the sensors
PO2 Cell Calibration menu options
O2 Sensor Polling

Oxygen cells how they sit in the Ouroboros rebreather
Cell block holder, Just disconect and the block comes free
Oxygen control system
(Note; Everything about the Ouroboros rebreather can be found in the comprehensive manual which accompanies every unit)

This can be a mechanical or an electronically controlled device. Which ever is chosen it is important that the control of the oxygen level is maintained within certain limits. If decompression is to be conducted using tables or a fixed PO2 dive computer, the limits must be accurately maintained. Rapid excursions to and from depth must also not generate excessively low or high PO2’s. Ideally any such limits reached should generate an alarm, as it cannot be relied upon that the diver will notice, especially if in a multi-tasking situation. Some rebreather designs, due to high flow rates within the oxygen circuit, can generate massive PO2 spikes sufficient to cause convulsions in a short space of time if left unchecked, this is an undesirable design feature should an addition valve fail. Within the CE EN14143 and most Navy tests is a PO2 tracking control test as well as a upper and lower limit test after rapid depth changes.

NOTES ON CELL HEALTH
Before any dive a comprehensive screen menu check system will occur. If any of the cells are not healthy then the screen on the right will displays levels below 20%.

This screen is only shown after an oxygen calibration, or if the cell health is low. It is also shown if the pre dive checks have been restarted from the option in the Setup screen and If cells are less than approximately 7mV in air or a wrong calibration gas has been used so that the cell mV appears to be less than 7mV in air., A warning is given not to dive.


Oxygen sensor isolation
The Ouroboros has the ability to automatically deselect faulty oxygen sensors. However there may be instances when you wish to override the computers decision. This can be done by manually turning off sensors in the O2 screen on page 61.
O2 sensor Screen
This screen allows the operator to view individual cell levels and status. The cells can also be disabled in the event of failure. When on the surface, the O2 sensors can be calibrated.


1;-From the O2 sensor screen, the operator can:
Enable/Disable sensor – Short press of both buttons
This allows the operator to toggle the sensor between an enabled (Y) and disabled (D) state. A disabled sensor will not be used in the sensor averaging. An enabled sensor will be used as long as its reading is in close proximity to the other enabled cells. The rebreather will not use cells that are out of range (displaying and N). It will also not use cells that have a dissimilar reading to the other cells. See o2 sensor polling on page 63 of our manual for more details on automatic disabling and averaging.



Above left; Dry Screen Above right; Diving Screen
 
2. Turn off rebreather oxygen solenoid control using all cells disabled (D)
If all 3 cells are turned to disabled (D) state, then the rebreather will stop controlling the oxygen solenoid. The diver will be prompted:
If the diver replies ‘yes’, then another warning screen appears and asks
:
If the diver again replies ‘yes’, then the solenoid control system is turned off and the decompression algorithm will follow an internally selected setpoint on the Primary display. The solenoid control system can be turned back on again (as will the cell tracking decompression) by changing one of the cells to ‘Y’ again.
ONLY ONE CELL WILL BE TURNED ON AGAIN THE NEXT TIME THE OUROBOROS IS TURNED OFF THEN BACK ON AGAIN. ENABLE CELLS AS APPROPRIATE.

3. Scroll down list – Short or long press of right button
This allows the operator to enter scroll down the list of cells
 

4. Enter CAL mode SURFACE ONLY – Long press of both buttons
This allows the operator to enter the o2 sensor calibrate screen. This is used to calibrate the po2 sensors and thus ensure that the rebreather is controlling as accurately as possible, and that the decompression calculations are using the appropriate po2 value.

O2 sensor Calibrate Screen
This screen allows the operator to calibrate the po2 sensors and thus ensuring that the rebreather is controlling PO2 as accurately as possible, and that the decompression calculations are using the appropriate po2 value.

NOTE: ANY SENSORS THAT THE OPERATOR THINKS ARE NOT WORKING CORRECTLY SHOULD BE REPLACED BEFORE ENTERING THIS SCREEN FROM THE O2 sensor Screen.

From the O2 sensor screen, the operator is prompted to perform ‘flush now’ with the Calibration Gas

When this has been done, a press of the right switch moves the operator to next screen. The operator is then prompted to ‘Vent’ the breathing loop momentarily by opening the mouthpiece and closing it again to equalize the pressure in the loop.

When this has been done, a press of the right switch moves the operator to next screen. The rebreather is then checking all the sensors for a stable reading. The rebreather will then automatically calibrate the sensors. Alternatively the user can use a short press of both switches to force the rebreather to calibrate. This is not advised, but in extreme cases can be used where the operator has no alternative but to use very unstable sensors.

When the calibration is complete, the operator is then taken to the first of the On screen Pre-dive checks. Now that the rebreather has an updated calibration, all pre dive functionality checks should be done to ensure the calibration is correct.
Calibration Gas
The calibration gas is setup from the Gas Adjust Screen. The calibration gas is the gas used in the rebreather O2 cylinder. This is normally 99% oxygen. However, should a slightly weaker O2 be used, then the CAL gas should be adjusted accordingly. Then the calibration of the O2 sensors will be adjusted accordingly.

Atmospheric Adjustment
Atmospheric adjustment is carried out automatically, during calibration, by multiplying the calibration gas O2 level by the current ambient pressure in Bar. This ensures the most appropriate PO2 is used by the rebreather.
Example: 98% O2 x 0.950 bar atmospheric = 0.931 Bar


Before deselecting a sensor, the PO2 should be confirmed by doing a diluent flush (with oxygen injection temporarily disabled). The PO2 of your selected diluent at the current depth is displayed on the dive Decompression Screen Summary on page 70 of our manual.

Calibration
It is not necessary to calibrate the unit prior to every dive but it is vital to do a pre-dive calibration check. When prompted to do so in the Pre-dive check screen (page 28), complete the following.
 

1. Make sure the diluent cylinder via the ADV cannot add gas into the loop (isolate the shut-off).
2. Evacuate the loop and close the mouthpiece.
3. Fill the loop with oxygen until the over pressure valve just releases.
4. Without adding exhaled air into the loop evacuate it again.
5. Fill and evacuate again twice, leaving the loop full of oxygen at the third fill.
6. Follow the on screen instructions.
7. Quickly open and close the mouthpiece to equalize the pressure in the loop with the surrounding ambient pressure.
8. Check the Backup display by pushing the top button to see the PO2 and re calibrate as required.
9. Check the Primary display and confirm the PO2 is correct on all 3 sensors to within 0.05 of the ambient PO2 with 100% oxygen.
10. If the PO2 is not correct, go to the GAS screen (page 58) and select Oxygen as the calibration gas. Now calibrate the unit.
11. Flush once more with oxygen and confirm the readings.
12. If the readings are still wrong you must abort the Pre-dive checks and open the Dome and expose the sensors to atmospheric air for 5 minutes. Now select the milli-volt function on the Backup display by pushing both buttons together. If any of the cells are below 7mVolts they need replacing.
13. Before replacing the cells remove and clean the jack plug connection to each cell. Check the ‘in-air’ reading again.
14. Replace and re calibrate as required.

Calibration at altitude is covered on page 64 of our official manual.

Oxygen sensor tests
If the cells become affected by moisture or are nearing the end of their life they may read incorrectly. Items to look for when a cell is suspect:
1. When flushing with oxygen, look for a cell reacting slowly. There may be moisture on the face. Remove the cell and dry it gently by applying tissue paper to the face. Do not apply pressure to the face.
2. Periodically check that the cells read greater than 7mV in air by pushing the bottom button on the Backup display. If a cell appears to go out of calibration quickly, flush the loop with oxygen. Calibrate the cells on the Primary and Backup displays. Leave the loop closed for one hour full of oxygen and then check the cells again on the Backup display without switching the main electronics on. They should all be within 0.1 PO2. Now flush with air, leaving the electronics off. The PO2 on the Backup should read between 0.23 and 0.19.

Calibrating the Backup Display
1. Ensure the battery is good.
2. When running the pre-dive sequence, once you have the loop flushed with oxygen, monitor the PO2 on the Backup Display and decide if any of the three displays need adjusting.
3. Turn the large blank cap on the rear of the Backup Display corresponding to the un-calibrated cell. If you run out of range on the control, try screwing it right in and start again. DO NOT SCREW THE CAP RIGHT OUT TO EXPOSE THE SEALING O RING AS THIS WILL FLOOD THE BACKUP DISPLAY, IF DIVED. If the backup will not calibrate then confirm the cell millivolt reading (by pushing the bottom button on the display) in air is above 7 millivolts and start the calibration routine again. Screw the cap right in and out again.
4. With the loop still full of oxygen, make sure the Backup Display oxygen reading is the same as the calibrated Primary Display readings.

5. If you completely undo the cap, clean and re-grease the waterproof cover O-ring and the internal O ring and refit the cap, being careful not to cross-thread the it.

PO2 Cell Calibration menu options

1. Flush Now
A short press of the right button confirms that the breathing loop has been properly flushed with the Calibration Gas.

2. Vent Now
A short press of the right button confirms that the breathing loop has been vented at the mouthpiece (quickly open and close) to ensure there is no overpressure in the breathing loop.

3. Calibrating..
After the ‘vent now’ operation is confirmed, the rebreather now monitors the active sensors, and performs the appropriate calibration calculations. This process can be stopped by a press of the left button. The process can also be manually confirmed by a short press of both buttons, in case the unit is taking an excessive time in getting a stable reading – e.g. because of water on the cells. However, normal practice would be to let the unit continue to calibrate automatically to ensure the best accuracy.


O2 Sensor Polling
The rebreather has a method of automatically removing O2 sensor cells from the PO2 averaging. This is based on a set of rules. If the operator considers these are not appropriate for a particular type of cell failure, then any individual cell can be turned off manually. See the O2 sensor Screen.

Rules:
1. If all cells have been disabled by the user the rebreather control system is turned off. See Dvo Screen and O2 sensor Screen for how to turn it back on.

2. If a single cell is below 0.15 bar or above 3.00bar, then it will be disabled, the system denotes this with a ‘N’ next to the cell.

3. If after item 2, all 3 cells are disabled for the same fault, then all cells will be re-enabled – this ensures that if the O2 is very high, or very low and all the cells agree, then the O2 is probably very high or low accordingly.

4. If all cells are enabled and have no faults, then each cell is checked to see how many other cells it is within 0.20bar of.
a. If all cells are within 0.20bar of each other, then all cells will be enabled.
b. If two cells are within 0.20bar of each other and one cell is not, then the cell that is not within 0.20bar of the others will be disabled.
c. If no cells are within 0.20bar of each other, then all cells will be kept enabled.

5. If all 3 cells are disabled with the same fault at this stage, then all will be re-enabled.

6. All enabled cells are then used in the PO2 averaging. Any cell disabled in these calculations will have a D or N shown against it in the O2 sensor Screen.

7. Examples:
a. Cell 1 = 0.5bar, cell 2 = 0.60bar, cell 3 = 0.70bar. All cells used (rule 4a)
b. Cell 1 = 0.3bar, cell 2 = 0.60bar, cell 3 = 0.70bar. Cells 2 and 3 only used (rule 4b)
c. Cell 1 = 0.3bar, cell 2 = 0.60bar, cell 3 = 0.14bar. Cell 1 and 2 only used (rule 2)
d. Cell 1 = 0.3bar, cell 2 = 0.60bar, cell 3 = 0.90bar. All cells used as no obvious fault in any single cell (rule 4c)


To Purchase cells please use the contact page to give us a call


top of page


 



Check out our rebreather Downloads >>>> Support | Technical | Training | History | Contact us | Gallery | Ouroboros
Content © VR Technology and associate photographers 2005 - 2008 All rights reserved
Photos on this site courtesy Leigh Bishop, Rob Smith, Jerome Meynie, Alexander Sotiriou